Back
Browse Products By:
APPLICATIONS
Back
Browse Products By Application
View All Applications
  • XRB160PN192

XRB160PN192

  • Integrated HV Supply, Filament Supply, X-Ray Tube, Beam Port and Control Electronics
  • Compact & Lightweight
  •  Universal Input, Power Factor Corrected
  • Can be Mounted in Any Physical Orientation
  • Analog Monitoring and Standard RS-232 Digital Interface
Get a Quote

XRB160PN192 Monoblock® Industrial X-Ray Generators

Spellman’s XRB160PN192 Monoblock® X-Ray source is designed for OEM applications powering its internal X-Ray tube up to 160kV at 192W. Features like universal input, small package size and a standard analog and RS-232 digital interface simplify integrating this Monoblock® into your X-Ray system. Standard models are available with fan shaped beam geometry. Proprietary emission control circuitry provides excellent regulation of X-Ray tube current, along with outstanding stability performance.

TYPICAL APPLICATIONS

X-Ray Scanning: Plating Measurement, Food Inspection, Fill Level Confirmation and Security Applications

  • Specifications
  • Models
  • Downloads

Specifications

  • X-Ray Characteristics:
    T
    ube Type:  Glass tube, Tungsten target, Be filter
    Focal Spot:  0.8mm x 0.8mm
    Beam Filter:  0.016˝ thick 6061 Al
    Beam Geometry: Asymmetrical fan 80° x 10° ±2° 
  • Input Voltage:
    100-240Vac ±10%, 50/60Hz, 5A maximum
  • X-Ray Tube Voltage:
    0.1mA to 1.2mA, 192W maximum over specified tube voltage range
  • X-Ray Tube Power:
    192W maximum continuous

Models

AC INPUT POWER J1 6 PIN CONNECTOR

PinSignalParameters
1LineLine
2RemovedN/C
3NeutralNeutral
Removed N/C 
Spare N/C 
Spare N/C 

 

TERFACE— J7 7 PIN MOLEX CONNECTOR 

PinSignalParameters
1Ex GateLow = X-Ray OFF, +12Vdc = X-Ray ON
2Signal GroundGround
3N/C No Connection
4kV Monitor0-9 Vdc = 0 to 100% rated output
5Signal GroundGround
6mA Monitor0 to 9Vdc = 0 to 100% rated output
FaultOpen collector, 35V @ 10mA max, High = No Fault

 

LED INDICATORS

PinSignalParameters
LED 1OVHigh kV occurs
LED 2UVLow kV occurs
LED 3UCLow mA occurs
LED 4OCHigh mA occurs
LED 5ARC FLTArc fault occurs
LED 6OTOver temperature occurs
LED 7X-RAY ONX-Rays are enabled
LED 8PWRPower is ON