XRBe80PN300 Monoblock® Series
- Compact & Lightweight
- Universal Input, Power Factor Corrected
- Can be Mounted in Any Physical Orientation
- Analog Monitoring Interface, Standard RS-232 Digital Interface and Ethernet
- Data Logging and Firmware Controlled X-Ray Tube Seasoning
80KV/300W Be Window Monoblock®
Spellman’s XRBe80PN300 beryllium window Monoblock® X-Ray sources are designed for OEM applications powering the internal Bipolar X-Ray tube at 80kV at a power level of 300W. Features such as universal input, compact package size and a standard RS-232 digital interface simplify integration of the XRBe80PN300 into your X-Ray system. It is available with a cone shaped beam geometry. Proprietary emission control circuitry provides excellent regulation of X-Ray tube current, along with outstanding stability performance.
TYPICAL APPLICATIONS
Food Inspection Systems, Fill Level Confirmation Systems, Security Scanning Systems, Industrial NDT Systems, Thickness/Plating Measurement Systems
- Specifications
- Downloads
Specifications
(Ref. 128132-001 REV. C)
- Options:
ST-TE, RA-SE, RA-TE
Cable Options (see page 3 of the data sheet) - X-Ray Characteristics:
- Focal Spot:
0.8mm (IEC 336) standard - Beam Filter:
Be 0.8mm - Beam Geometry:
- Cone:
Standard. The beam angular coverage will be 40° with the beam plane perpendicular to the X-Ray tube axis - Input Voltage:
100- 240Vac, ±10%, 50/60 Hertz, .98 power factor - Input Current:
300W @ 6A - X-Ray Tube Voltage:
Maximum 80kV - X-Ray Tube Current:
Maximum 5mA - X-Ray Tube Power:
300W - Voltage Regulation:
- Line:
Line:±0.05% of maximum output voltage over a ±10% change of nominal input line voltage - Load:
±0.1% of maximum rated voltage for 150uA to full rated load change - Voltage Accuracy:
Voltage measured across the X-Ray tube is within ±2% of the programmed value - Voltage Risetime:
Standard ramp time shall be <500ms from 10% to 90% of maximum rated output voltage - Voltage Ripple:
0.5% peak to peak of maximum voltage forfrequencies ≤1kHz - X-Ray Tube Voltage:
Maximum 80kV
Emission Current Parameters - Current Regulation:
- Line:
±0.05% of rated output current over a ±10% change of nominal input line voltage - Load:
±0.1% of rated output current for a change from 50% to 100% of rated output voltage - Current Risetime:
Standard ramp time shall be <500ms from 10% to 90% of maximum rated current - Arc Intervention:
4 arcs in 10 seconds with a 100ms quench/100ms re-ramp = Shutdown - Filament Configuration:
Internal floating AC filament drive with closed loop filament emission control circuitry