80kV, 100W X-Ray Source
Spellman’s XRB80N100 Monoblock® X-Ray source is designed for OEM applications powering its internal X-Ray tube up to 80kV at 100W. Features like universal input, small package size and a standard analog and RS-232 digital interface simplify integrating this unit into your X-Ray system. The XRB80N100 is available either with fan shaped (standard) or (optional) cone shaped beam geometries. Proprietary emission control circuitry provides excellent regulation of X-Ray tube current, along with outstanding stability performance.
Typical XRB80N100 Monoblock® Integrated X-Ray Source applications:
- X-Ray Scanning: Bone Densitometry
- Thickness Measurement
- Food Inspection
- Fill Level Confirmation
- Parcel Inspection
- Specifications
- Downloads
Specifications
(Ref. 128065-001 REV. Y)
- X-Ray Characteristics:
- Focal Spot:
0.5mm (IEC 336) - Beam Filter:
- Ultem:
3.30mm ±0.15mm - Oil:
8mm ±0.1mm - Glass:
1.8mm ±0.25mm - BE:
0.8mm - Beam Geometry:
- Fan:
Standard. The beam angular coverage will be 75° with the beam plane perpendicular to the X-Ray tube axis and 13° wide. - Cone:
Optional. 25° cone beam - Input Voltage:
Power factor corrected input 0.98, 100-240Vac ±10%
50/60Hz, 2A, maximum - X-Ray Tube Voltage:
Nominal X-Ray tube voltage is adjustable up to 80kV - X-Ray Tube Current:
150uA to 1.25mA over specified tube voltage range - X-Ray Tube Power:
100W, maximum continuous - Voltage Regulation:
- Line:
±0.05% of maximum output voltage over a ±10% change of nominal input line voltage - Load:
±0.1% of maximum rated voltage for 150uA to 1.25mA load change - Voltage Accuracy:
Voltage measured across the X-Ray tube is within ±2% of the programmed value - Voltage Risetime:
- Standard:
Ramp time shall be 500ms from 10% to
90% of maximum rated output voltage - Optional:
5 seconds. Specify at time of order - Voltage Overshoot:
5% of maximum voltage, to return within 2.5% of
maximum voltage in less than 100ms - Voltage Ripple:
1% peak to peak of maximum voltage for
frequencies
≤
1kHz
Emission Current Parameters - Current Regulation:
- Line:
±0.05% of rated output current over a ±10% change of nominal input line voltage - Load:
±0.1% of rated output current for a change from 50% to 100% of rated output voltage - Current Accuracy:
Current measured through the X-Ray tube is within ±2% of the programmed value - Current Risetime:
- Standard:
Ramp time shall be 500ms from 10% to 90% of maximum rated current - Optional:
5 seconds. Specify at time of order - Arc Intervention:
3 arcs in 10 seconds with a 200ms quench = Shutdown - Filament Configuration: