XRB100N100 Monoblock® Series
- Integrated HV Supply, Filament Supply, X-Ray Tube, Beam Port and Control Electronics
- Compact & Lightweight
- Universal input, Power Factor Corrected with Internal EMI Filter
- Can be Mounted in Any Physical Orientation
- Analog Control Interface and Standard RS-232 Digital Interface
Spellman’s XRB100N100 Monoblock® X-Ray source is designed for OEM applications powering its internal X-Ray tube up to 100kV at 100W. Features like universal input, small package size and a standard analog and RS-232 digital interface simplify integrating this Monoblock® into your X-Ray system. Proprietary emission control circuitry provides excellent regulation of X-Ray tube current, along with outstanding stability performance.
Typical applications:
- X-Ray Scanning: Food Inspection
- Fill Level Confirmation
- Security Applications
- Specifications
- Downloads
Specifications
(Ref. 128072-001 REV. L)
- X-Ray Characteristics:
- Tube Type:
Stationary anode, tungsten target - Focal Spot:
0.5mm (IEC 336) - Beam Filter:
- Lexan:
3.2mm - Oil:
10mm ±0.1mm - Glass:
1.8mm max - Beam Geometry:
Symmetrical fan 74° x 10° ±1° - Input Voltage:
Power factor corrected input 0.98, 100-240Vac ±10%
50/60Hz, 2A, maximum - X-Ray Tube Voltage:
Nominal X-Ray tube voltage is adjustable between
40kV to 100kV - X-Ray Tube Current:
100uA to 1mA over specified tube voltage range - X-Ray Tube Power:
100W, maximum continuous - Voltage Regulation:
- Line:
±0.1% of maximum output voltage over a
±10% change of nominal input line voltage - Load:
±0.1% of maximum rated voltage for 100uA to 1mA load change - Voltage Accuracy:
Voltage measured across the X-Ray tube is within ±2% of the programmed value - Voltage Risetime:
Ramp time shall be 1 second from 10% to 90% of maximum rated output voltage - Voltage Overshoot:
5% of maximum voltage, to return within 2.5% of maximum voltage in less than 50ms - Voltage Ripple:
0.5% peak to peak of maximum voltage for frequencies
≤
1kHz - Emission Current Parameters Current Regulation:
- Line:
±0.5% of rated output current over a
±10% change of nominal input line voltage - Load:
±0.5% of rated output current for a change from 50% to 100% of rated output voltage - Current Accuracy:
Current measured through the X-Ray tube is within ±1% of the programmed value - Current Risetime:
Ramp time shall be 1 second from 10% to 90% of maximum rated current - Arc Intervention:
3 arcs in 10 seconds with a 200ms quench = Shutdown - Filament Configuration:
Internal high frequency AC filament drive with closed loop filament emission control