XRB011 Monoblock® Series
- Integrated HV Supply, Filament Supply, X-Ray Tube, Beam Port and Control Electronics
- Compact & Lightweight
- Can be Mounted in Any Physical Orientation
- Analog or Digital Control Interface
Certifications & Compliance
Spellman’s XRB011 Series of Monoblock® X-Ray sources are designed for OEM applications powering its internal X-Ray tube up to 80kV at 20W and 80kV at 50W. Features like 24Vdc input voltage, small package size, standard analog interface and RS-232/Ethernet digital interface simplify integrating the XRB011 into your X-Ray system. Proprietary emission control circuitry provides excellent regulation of X-Ray tube current, along with outstanding stability and performance.
Typical applications:
- Medical X-Ray: Fluoroscopy and Radiography for Extremities, Specimen Radiography.
- Pulsed Fluoroscopy (contact Spellman sales)
- Industrial X-Ray: Component inspection and Non-Destructive Testing
- Specifications
- How to Order
- Downloads
Specifications
(Ref. 128090-001 REV. AF)
- X-Ray Characteristics:
- Tube Type:
Micro focus tube - Focal Spot:
33µm Nominal, 50µm max. (IEC 336) - Beam Filter:
Ultem 0.060. (1.5mm)
Oil 0.175. (4.4mm) - Beam Geometry:
Symmetrical 40° cone - Input Voltage:
- 20W:
24Vdc ±1V @ 2.5A - 50W:
24Vdc ±1V @ 4A - X-Ray Tube Voltage:
Nominal X-Ray tube voltage is adjustable between 35kV to 80kV - X-Ray Tube Current:
- 20W:
0-250µA over specified tube voltage range - 50W:
0-700µA over specified tube voltage range - X-Ray Tube Power:
20/50W, maximum continuous - Voltage Regulation:
- Line:
±0.5% for a ±1V change of nominal input line voltage - Load:
±0.1% for a load change of 25µA to maximum rated current - Voltage Accuracy:
Voltage measured across the X-Ray tube is within ±1% of the programmed value - Voltage Risetime:
Ramp time shall be =250ms from 10% to 90% of maximum rated output voltage - Voltage Temperature Coefficient:
≤100ppm/°C - Over Temperature Fault:
Indicates that the internal oil temperature has exceeded 65° C. The high voltage output will be disabled. Toggling the X-Ray ON Command OFF and ON will reset the fault. - Over Voltage Fault:
An overvoltage (OV) fault is detected when the output voltage exceeds 82kV. The high voltage output will be disabled. Toggling the X-Ray ON Command OFF and ON will reset the fault. - Voltage Ripple:
1% peak to peak - Current Regulation:
- Line:
±0.5% for a ±1V change of nominal input line voltage - Load:
±0.5% for a voltage change of 35kV to 80kV - Current Accuracy:
Current measured through the X-Ray tube is within ±2.5% of the programmed value - Over Current Fault:
An overcurrent (OC) fault is detected when the emission current exceeds 275µA (20W model) and 710µA (50W model). Toggling the X-Ray ON Command OFF and ON will reset the fault. - Arc Intervention:
One arc fault. The high voltage output will be disabled. Toggling the X-Ray ON command OFF and ON will reset the fault. - Filament Configuration:
Internal high frequency AC filament drive with closed loop filament emission control
How to Order
ORDERING INFORMATION
Medical Applications:
XRB011-80PN20 80kV, 250uA, 20W, Analog Interface, RS-232
XRB011-80PN20E 80kV, 250uA, 20W, Analog Interface, RS-232, Ethernet
XRB011-80PN20A 80kV, 250uA, 20W, Analog Interface
Non-Medical Applications:
XRB011-80PN20/CE 80kV, 250uA, 20W, Analog Interface, RS-232, CE
XRB011-80PN20E/CE 80kV, 250uA, 20W, Analog Interface, RS-232, Ethernet, CE
XRB011-80PN20A/CE 80kV, 250uA, 20W, Analog Interface, CE
50W Model
ORDERING INFORMATION
Medical Applications:
XRB011-80PN50 80kV, 700uA, 50W, Analog Interface, RS-232
XRB011-80PN50E 80kV, 700uA, 50W, Analog Interface, RS-232, Ethernet
XRB011-80PN50A 80kV, 700uA, 50W, Analog Interface
Non-Medical Applications:
XRB011-80PN50/CE 80kV, 700uA, 50W, Analog Interface, RS232, CE
XRB011-80PN50E/CE 80kV, 700uA, 50W, Analog Interface, RS232, Ethernet, CE
XRB011-80PN50A/CE 80kV, 700uA, 50W, Analog Interface, CE