PDM90PN900 Monoblock®

  • X-Ray Source with Integrated Filament, Tube and Beam Port
  • Separate Control Electronics Box
  • Compact & Lightweight
  • Analog or Digital Control Interface

Certifications & Compliance

Get a Quote

Panoramic Dental X-Ray Source

Spellman’s new Panoramic Dental Monoblock® consists of an integrated X-Ray tube, dual-output high voltage power supply, and filament supply with control circuitry. The PDM90PN900 is designed for extra oral dental X-Ray applications including CT and Panoramic X-Ray. Features like small package size, standard analog and RS-232digital interface simplify integrating this Monoblock® into your X-Ray system. Proprietary emission control circuitry provides excellent regulation of X-Ray tube current, along with outstanding stability and performance.

  • Specifications
  • Downloads

Specifications

(Ref. 128109-001 REV. D)

  • X-Ray Tube Characteristics:

    Target Angle:

    5 degrees

    Focal Spot:

    0.5mm nominal

    Beam Filtration:

    3.0mm of Al equivalent
  • X-Ray Tube Voltage:
    Nominal X-Ray tube voltage is adjustable between
    50kV to 90kV
  • Voltage Regulation:

    Line:

    ±0.5% for a ±1V change of nominal input line voltage

    Load:

    ±0.1% for a load change of 25µA to maximum rated current
  • X-Ray Tube Current:
    1mA to 10mA over specified tube voltage range
  • Current Regulation:

    Line:

    ±0.5% for a ±1V change of nominal input line voltage

    Load:

    ±0.5% for a voltage change of 35kV to 80kV
  • X-Ray Tube Power:
    900 watts peak power
  • Duty Cycle:

    CT (Pulsed):

    ≤ 45 seconds scan time cycle at up to 40 pulses per second available

    Panoramic (Continuous):

    Maximum scan time = 30s, with 60s off-time. Five consecutive scans
  • Input Voltage:
    100-240Vac ±10% 50/60 Hz, 10 amps RMS maximum
  • Interface:
    The RS-232 serial communications interface will be used to program and monitor output voltage and current, control various functions and report status and faults.
  • Digital Interface Connector:

    RS-232:

    9 pin D connector, male
  • Operating Temperature:
    0°C to +40°C
  • Storage Temperature:
    -20°C to +70°C
  • Humidity:
    5% to 95% relative humidity, non-condensing
  • Cooling:

    Tank:

    Convection

    Controller:

    Forced air via provided fan
  • Dimensions:

    X-Ray Tank:

    9.7"W X 7.7"H X 4.7"D (247mm X 195mm X 119mm)

    Inverter/Controller:

    7.10"W X 9.80"H X 3.60"D (180mm X 250mm X 92mm)
  • Weight:

    X-Ray Tank:

    17lbs (7.7kg)

    Inverter/Controller:

    6.5lbs (3kg)
  • X-Ray Leakage:
    Less than 100mR/hour (or <1mGy/hr) @ 1meter from the Monoblock® surface.
  • Regulatory Approvals:
    Compliant to EMC 60601-1-2 (external EMC filter and shielding required). UL/CUL recognized file E242584.

Typical applications:

  • Dental X-Ray: Panoramic and CT Imaging
Service Solution Brochure

Spellman Repair Centers, Extended Warranty Programs, Preventative Maintenance Programs, Calibration Services and more.

Download
2026 High Voltage Reference Manual

We compiled all of our useful FAQ’s, application notes, articles, technical papers and three glossaries into a single handy reference guide.

Download