  • ±10kV with 25mS Polarity Reversing Speed
  • Integrated Electron Multiplier Supply
  • Precision Analog Voltage Control
  • High Stability, Low Ripple and Noise
  • High Voltage Inhibit Control
  • RoHS Compliant 

Certifications & Compliance

Get a Quote

High Performance High Voltage Power Supply

Spellman’s RoHS compliant MX10 Plus is a well-regulated high performance fast reversible dynode supply featuring a 25ms “hot switchable” polarity reversing capability with an integrated -2.3kV electron multiplier supply.

The MX10 Plus’s low ripple specification is typical of the topologies that make Spellman High Voltage your ideal choice for mass spectrometry applications; especially security detection systems, dynodes, sample ionisation as well as capillary electrophoresis and electrostatic printing applications. The MX10 Plus has been designed espe­cially for dynode detector applications.

The MX10 Plus can be easily tailored to an OEM’s requirement, such as improved ripple performance, or different voltage and/or current capabilities.


  • Specifications
  • How to Order
  • Downloads

Specifications

  • Input Voltage:
    +15Vdc, ±.75Vdc
  • Input Current:
    ≤500mA nominal continuous
    <2A during reversing
  • Temperature Coefficient: ≤100ppm per degree C
  • Stability: (constant operating conditions) ≤300ppm per hour after 1 hour warm up
  • Regulatory Approvals:
    Compliant to EEC EMC Directive. Compliant to EEC Low Voltage Directive.
    RoHS Compliant.
  • DYNODE SPECIFICATIONS
    Output Voltage:

    ±10kV
    Output Current:

    10µA
    Output Polarity:

    Remotely reversible via TTL logic signal
    Switching Speed:
    25ms to settle 90% into 50pF load
    Voltage Regulation:
    Line: ≤0.02% for a 1.5V input voltage change
    Ripple:
    ≤10 Volts p-p
  • ELECTRON MULTIPLIER SPECIFICATIONS
    Output Voltage:
    Fixed:
    -2.3kV
    Output Polarity:
    Negative
    Output Current:
    ≤ 230µA
    Voltage Regulation:
    Line: ≤0.02% for a 1.5V input voltage change
    Load:
    <5V for for no load to 22M Ohms load change
    Ripple:
    ≤200mV p-p @ 2.3kV into 22M Ohm load Output Rise Time: 10ms Output Fall Time: 10ms
    Output Rise Time: 10ms Output
    Fall Time: 10ms
  • Dimensions:
    2.00˝ H X 5.30˝ W X 8.00˝ D (50.8mm X 134.6mm X 203mm)
  • Weight:
    Approximately 3.3 pounds (1.5kg)

How to Order

How to Order:
Standard: PART NO.:MXP10PN15

Typical Applications:

  • Mass Spectrometry
  • Sample Ionisation
  • Capillary Electrophoresis
Service Solution Brochure

Spellman Repair Centers, Extended Warranty Programs, Preventative Maintenance Programs, Calibration Services and more.

Download
2026 High Voltage Reference Manual

We compiled all of our useful FAQ’s, application notes, articles, technical papers and three glossaries into a single handy reference guide.

Download