뒤로
다음 기준으로 제품 찾아보기:
응用 분야
뒤로
응용 분야별 제품 보기
모든 응용 분야 보기

XRBe80PN300 Monoblock® Series

  • Compact & Lightweight
  • Universal Input, Power Factor Corrected
  • Can be Mounted in Any Physical Orientation
  • Analog Monitoring Interface, Standard RS-232 Digital Interface and Ethernet
  • Data Logging and Firmware Controlled X-Ray Tube Seasoning

*Note: All specifications are subject to change without notice. Please consult the English PDF version of this datasheet for the most up-to-date revision.

견적 받기

80KV/300W Be Window Monoblock®

Spellman’s XRBe80PN300 beryllium window Monoblock® X-Ray sources are designed for OEM applications powering the internal Bipolar X-Ray tube at 80kV at a power level of 300W. Features such as universal input, compact package size and a standard RS-232 digital interface simplify integration of the XRBe80PN300 into your X-Ray system. It is available with a cone shaped beam geometry. Proprietary emission control circuitry provides excellent regulation of X-Ray tube current, along with outstanding stability performance.

TYPICAL APPLICATIONS

Food Inspection Systems, Fill Level Confirmation Systems, Security Scanning Systems, Industrial NDT Systems, Thickness/Plating Measurement Systems

  • 사양
  • 다운로드

사양

(Ref. 128132-001 REV. C)

  • Options:
    ST-TE, RA-SE, RA-TE
    Cable Options (see page 3 of the data sheet)
  • X-Ray Characteristics:
  • Focal Spot:
    0.8mm (IEC 336) standard
  • Beam Filter:
    Be 0.8mm
  • Beam Geometry:
  • Cone:
    Standard. The beam angular coverage will be 40° with the beam plane perpendicular to the X-Ray tube axis
  • Input Voltage:
    100- 240Vac, ±10%, 50/60 Hertz, .98 power factor
  • Input Current:
    300W @ 6A
  • X-Ray Tube Voltage:
    Maximum 80kV
  • X-Ray Tube Current:
    Maximum 5mA
  • X-Ray Tube Power:
    300W
  • Voltage Regulation:
  • Line:
    Line:±0.05% of maximum output voltage over a ±10% change of nominal input line voltage
  • Load:
    ±0.1% of maximum rated voltage for 150uA to full rated load change
  • Voltage Accuracy:
    Voltage measured across the X-Ray tube is within ±2% of the programmed value
  • Voltage Risetime:
    Standard ramp time shall be <500ms from 10% to 90% of maximum rated output voltage
  • Voltage Ripple:
    0.5% peak to peak of maximum voltage forfrequencies ≤1kHz
  • X-Ray Tube Voltage:
    Maximum 80kV
    Emission Current Parameters
  • Current Regulation:
  • Line:
    ±0.05% of rated output current over a ±10% change of nominal input line voltage
  • Load:
    ±0.1% of rated output current for a change from 50% to 100% of rated output voltage
  • Current Risetime:
    Standard ramp time shall be <500ms from 10% to 90% of maximum rated current
  • Arc Intervention:
    4 arcs in 10 seconds with a 100ms quench/100ms re-ramp = Shutdown
  • Filament Configuration:
    Internal floating AC filament drive with closed loop filament emission control circuitry