XRB180PN200 Monoblock®

  • 180kV and 200W
  • X-Ray Tube Current: 0.5mA to 1.1mA
  • Focal Spot 0.8 x 0.8mm nominal (IEC60336)
  • Power factor corrected input to 0.98
  • 100Vac to 240Vac, ±10%

*Note: All specifications are subject to change without notice. Please consult the English PDF version of this datasheet for the most up-to-date revision.

견적 받기

Spellman’s XRB180PN200 is an integrated X-Ray source operating up to 180kV and 200W, providing OEM users a compact plug-and-play Monoblock for critical inspection and screening applications. The unit incorporates a 25° angle stationary anode X-Ray tube offering a fan beam of 80°x 4°. Proprietary emission control circuitry provides excellent regulation of X-Ray tube current, with industry leading dose stability and image quality. This compact model comes with standard analog and RS-232 digital control. Spellman can provide customized versions of this platform for specific OEM system requirements.

TYPICAL APPLICATIONS

Aviation Security Screening: Checked Baggage/EDS, Checkpoint Vehicle Inspection, General NDT, Food Inspection

  • 사양
  • 다운로드

사양

(Ref. 128085-001 REV. A)

  • X-Ray Characteristics:
  • Tube Type:
    Stationary Anode, tungsten target
  • Focal Spot:
    0.8 x 0.8mm nominal (IEC60336)
  • Beam Filter:
    Consists of oil and the curved Ultem 0.079'' +/-0.01''. Inherent filtration in the X-Ray Tube is 1mm Be, 1.5mm +/-10% glass.
  • Beam Geometry:
    Fan of 80° X 4°, ±1%
  • Anode Angle:
    25°
  • Input Voltage:
    Power factor corrected input to 0.98. 100Vac to 240Vac, ±10%, 50/60 Hertz. 5A, maximum
  • X-Ray Tube Voltage:
    Nominal X-Ray tube voltage is adjustable between 90kV to 180kV.
  • Voltage Accuracy:
    The high voltage measured at the X-ray tube will be within ±1% of the selected value.
  • Voltage Ripple:
    Ripple will be ≤0.2% of maximum rated voltage for frequencies ≤1kHz
  • Voltage Regulation:
    < ±0.1% for ±10% of nominal input line change
    < ±0.1% for 0.5mA to 1.11mA load change
  • Voltage Overshoot:
    kV overshoot will return within 5% of full voltage in less than 10ms
  • Risetime:
    The voltage and current risetime is controlled by a ramping circuit. Ramp time is less than 0.5 second from 10% to 90% of output voltage and current.
  • X-Ray Tube Current:
    0.5mA to 1.1mA @ 180kV, 200 Watts maximum
  • Current Accuracy:
    <±1% of the selected value.
  • Current Regulation:
    <0.5% at 90-180kV, 0.5mA to 1.1mA
  • Arc Intervention:
    Unit will detect a single arc but HV will not shut down. If multiple arcs occur (4 in 10 seconds) then the unit will shut down.
  • Filament Configuration:
    Internal high frequency AC filament drive operated in current mode with closed loop filament emission control.