XRB150PN600 Monoblock®

  • Integrated HV Supply, Filament Supply, X-Ray Tube, Beam Port and Control Electronics
  • Compact & Lightweight
  • Power Factor Corrected
  • Can be Mounted in Any Physical Orientation
  • Analog Monitoring and Standard RS-232 Digital Interface
견적 받기

150kV, 600W X-Ray Source

Spellman’s XRB150PN600 Monoblock® X-Ray source is designed for OEM applications powering its internalX-Ray tube up to 150kV at 600W. Features like power factor correction, small package size and a standard analog and RS-232 digital interface simplify integrating this unit into your X-Ray system. Standard models are available with cone shaped beam geometry. Proprietary emission control circuitry provides excellent regulation of X-Ray tube current, along with outstanding stability performance.

Typical applications:

  • X-Ray Scanning: Bone Densitometry
  • Food Inspection
  • Security
  • 사양
  • 다운로드

사양

(Ref. 128079-001 REV. G)

  • X-Ray Characteristics:
  • Tube Type:
    Glass tube, Tungsten target, Be filter
  • Focal Spot:
    0.5mm x 0.5mm
    (IEC 336)
  • Beam Filter:
    0.06. Ultem
  • Beam Geometry:
    C
    one up to 18° ±1°
  • Input Voltage:
    Vac, 50/60Hz, 6.5A maximum 200-240 +/-10%
  • X-Ray Tube Voltage:
    Nominal X-Ray tube voltage is adjustable between 70kV to 150kV
  • X-Ray Tube Current:
    1.0mA to 4.0mA over specified tube voltage range
  • X-Ray Tube Power:
  • 160W continuous, 600W peak Duty Cycle:
    30 seconds on, 300 seconds off @ 600W peak
  • Voltage Regulation:
  • Line:
    ±0.1% for a ±10% input line change of 180 to 264Vac
  • Load:
    ±0.1% for a 1.0mA to 4.0mA load change. 600W maximum
  • Voltage Accuracy:
    Voltage measured across the X-Ray tube is within ±2% of the programmed value
  • Voltage Risetime:
    ±1% in less than 300ms
  • Voltage Overshoot:
    ±10% during 300ms risetime
  • Voltage Ripple:
    1% rms of rated voltage @ 10Hz to 1MHz
  • Current Regulation:
  • Line:
    ±0.1% from 180-264Vac
  • Load:
    ±0.5% @ 70-150kV, 1.0mA to 4.0mA
  • Current Accuracy:
    Current measured through the X-Ray tube is within ±2% of the programmed value
  • Current Risetime:
    ±1% in less than 300ms
  • Arc Intervention:
    4 arcs in 10 seconds with a 200ms quench = Shutdown
  • Filament Configuration:
    Internal high frequency AC filament drive with closed loop filament emission control