뒤로
다음 기준으로 제품 찾아보기:
응用 분야
뒤로
응용 분야별 제품 보기
모든 응용 분야 보기

PDM90PN900 Monoblock®

  • X-Ray Source with Integrated Filament, Tube and Beam Port
  • Separate Control Electronics Box
  • Compact & Lightweight
  • Analog or Digital Control Interface

Certifications & Compliance

견적 받기

Spellman’s new Panoramic Dental Monoblock® consists of an integrated X-Ray tube, dual-output high voltage power supply, and filament supply with control circuitry. The PDM90PN900 is designed for extra oral dental X-Ray applications including CT and Panoramic X-Ray. Features like small package size, standard analog and RS-232digital interface simplify integrating this Monoblock® into your X-Ray system. Proprietary emission control circuitry provides excellent regulation of X-Ray tube current, along with outstanding stability and performance.

Typical applications:

  • Dental X-Ray: Panoramic and CT Imaging
  • 사양
  • 다운로드

사양

(Ref. 128109-001 REV. D)

  • X-Ray Tube Characteristics:

    Target Angle:

    5 degrees

    Focal Spot:

    0.5mm nominal

    Beam Filtration:

    3.0mm of Al equivalent
  • X-Ray Tube Voltage:
    Nominal X-Ray tube voltage is adjustable between
    50kV to 90kV
  • Voltage Regulation:

    Line:

    ±0.5% for a ±1V change of nominal input line voltage

    Load:

    ±0.1% for a load change of 25µA to maximum rated current
  • X-Ray Tube Current:
    1mA to 10mA over specified tube voltage range
  • Current Regulation:

    Line:

    ±0.5% for a ±1V change of nominal input line voltage

    Load:

    ±0.5% for a voltage change of 35kV to 80kV
  • X-Ray Tube Power:
    900 watts peak power
  • Duty Cycle:

    CT (Pulsed):

    ≤ 45 seconds scan time cycle at up to 40 pulses per second available

    Panoramic (Continuous):

    Maximum scan time = 30s, with 60s off-time. Five consecutive scans
  • Input Voltage:
    100-240Vac ±10% 50/60 Hz, 10 amps RMS maximum
  • Interface:
    The RS-232 serial communications interface will be used to program and monitor output voltage and current, control various functions and report status and faults.
  • Digital Interface Connector:

    RS-232:

    9 pin D connector, male
  • Operating Temperature:
    0°C to +40°C
  • Storage Temperature:
    -20°C to +70°C
  • Humidity:
    5% to 95% relative humidity, non-condensing
  • Cooling:

    Tank:

    Convection

    Controller:

    Forced air via provided fan
  • Dimensions:

    X-Ray Tank:

    9.7"W X 7.7"H X 4.7"D (247mm X 195mm X 119mm)

    Inverter/Controller:

    7.10"W X 9.80"H X 3.60"D (180mm X 250mm X 92mm)
  • Weight:

    X-Ray Tank:

    17lbs (7.7kg)

    Inverter/Controller:

    6.5lbs (3kg)
  • X-Ray Leakage:
    Less than 100mR/hour (or <1mGy/hr) @ 1meter from the Monoblock® surface.
  • Regulatory Approvals:
    Compliant to EMC 60601-1-2 (external EMC filter and shielding required). UL/CUL recognized file E242584.