MFX Series
- 50kV with 50 or 75 Watt Max
- 65kV with 65 Watt Max
- Floating Integrated Filament Supply
- Local and Remote Emission Control
- Optional Digital Interface
Certifications & Compliance
The MFX Series is the result of Spellman’s exceptional high voltage packaging and surface mount fabrication techniques, coupled with its proprietary encapsulation technology producing this ultra compact-sized OEM X-Ray generator module.
The MFX Series is designed to power floating filament X-Ray tubes from various manufacturers. It features a 0 to 50kV (or 0 to 65kV) high voltage output, and up to 2mA of emission current limited to 50, 65 or 75 Watts, operating from a +24Vdc input. The MFX utilizes a closed loop filamentry beam control circuit to provide a highly regulated beam current. The floating ac filament supply operates between 0.3 and 4 amps. Offering tight regulation, high stability and low ripple, the MFX provides users both local and remote analog control to set beam voltage, emission current and filament current limit. An optional USB, RS-232 and Ethernet interface is available.
OPTIONS
SIC Standard Interface Controller
(Ethernet, USB & RS-232)
5VPM 0 to 5 Volt Programming and Monitor Scaling
- 사양
- 주문 방법
사양
(Ref. 128101-001 REV. K)
- Input:
+24Vdc ±10%, 5.0A maximum for either 50 Watts or 75 Watts.
+24Vdc ±1V, 5.0A maximum for 65kV/65W units. - Output:
0 to 50kV at 0 to 2mA, limited to a maximum of 50 watts or 75 Watts.
0-65kV at 2mA limited to 65 Watts. Negative output - High Frequency AC Filament Supply:
Current: 0-4A, adjustable limit
Voltage: 5.0 volt limit - Voltage Regulation:Load: 0.01% of output voltage no load to full load.
Line: ±0.01% for ±10% change in input voltage.
Current Regulation:
Load: 0.01% of output current from 0 to rated voltage.
Line: ±0.01% for ±10% change in input voltage. - Ripple:
0.1% p-p of maximum rated output voltage, 50kV units (50/75W)
0.1% rms maximum rated output voltage, 65kV units (65W) - Temperature Coefficient:
0.01% per °C, voltage and current. - Stability: 0.05% per 8 hours after 1/2 hour warm-up.
- Weight: 7.0 lbs. (3.2kg)
- Dimensions:
50kV Unit:
7.00”H x 2.87”W x 8”D (177.80mm x 72.90mm x 203.25mm)
65kV Unit:
7.00”H x 2.87”W x 9”D (177.80mm x 72.90mm x 228.65mm)
SIC Option:
7.75”H x 2.87”W x 8”D/9”D (196.85mm x 72.90mm x 203.25mm/228.65mm) - Weight:
7.0 lbs. (3.2kg) - Regulatory Approvals:
RoHS compliant.
주문 방법
| How to Order: |
|---|
|
Sample model number: |
| Options are added to the model number as follows: MFX50N75/SIC |
Typical Applications:
- Analytical X-Ray
- X-Ray diffraction
- Thickness Gauging
스펠만 고전압 뉴스레터를 구독하세요
분기마다 발행되는 스펠만 고전압 뉴스레터를 구독하세요. 고전압전원장치에 대한 기술 정보와 최신 업계동향을 전달하여 드립니다.