ML1350
- Quad Output High Voltage Power Supply
- Low Cost, Compact Air Insulated Design
- Remote Enable Control Provided
- UL Approved to UL61010-1
*Note: All specifications are subject to change without notice. Please consult the English PDF version of this datasheet for the most up-to-date revision.
Mass Spectrometry High Voltage Power Supply
Spellman’s ML1350 power supply module has been designed specifically to drive quadrupoles used in mass spectrometry. This compact, low cost, SMT based high performance module is printed circuit board mountable. This quad output supply is ideal for quadrupole drivers and electrostatic lenses.
This voltage regulated, current limited, fixed quad output unit provides up to 15mA of load current from each output. The ML1350 is fully protected against arc and short circuit conditions. The grounded metal case provides both shielding and heat sinking functions. An Enable feature is provided, allowing simple remote operation of the supply. The ML1350 is CE and UL approved.
Typical applications:
- Quadrupole HVPS
- Electrostatic Lenses
Specifications
(Ref. 128037-001 REV. J)
Input Voltage:
+24 Vdc, ±1.2Vdc
Input Current:
≤ 3.0 amps
Output Voltage:
Output 1-Positive:
+245 volts, fixed, accuracy <±10%
Output 2-Negative:
-245 volts, fixed, accuracy <±10%
Output 3-Positive:
+1350 volts, fixed, accuracy <±7%
Output 4-Negative:
-1350 volts, fixed, accuracy <±7%
Output Current:
15mA maximum for each output
Line Regulation: (typical)
±1 volt all outputs
Load Regulation: (typical)
±3% all outputs
Output Current Limit:
An auto-recovering short circuit fold back limit is employed. Fully arc protected, capable of 10 arcs in 5 seconds.
Ripple:
≤ 0.1% p-p of full rated output voltage
Stability:
≤ 0.25% per hour, constant operating conditions after 1 hour warm up.
Under Voltage Shutdown:
The power supply will shut down when an input under voltage condition is detected. When the input voltage is restored above 11.8 volts, operating the enable pin will reset this fault.
Temperature Coefficient:
≤ 200ppm per degree C
Environmental:
Temperature Range:
Operating: 0°C to 50°C
Storage: -35°C to 85°C
Humidity:
10% to 90% RH, non-condensing
Cooling:
Unit must be mounted in free air, in any position with the exception of inverted (pins up). Forced air cooling is recommended.
Dimensions:
0.984. H X 4.331. W X 3.150. D (25mm x 110mm x 80mm) Width does not include mounting tab
Weight:
7.27 oz. (206g)
Regulatory Approvals:
Compliant to EEC EMC Directive. Compliant to EEC Low Voltage Directive.
Tables & Diagrams
DIMENSIONS: in.[mm]
TOP VIEW
FRONT VIEW