Back
Browse Products By:
APPLICATIONS
Back
Browse Products By Application
View All Applications

CZE1000R

  • Ideal for Electrospinning
  • 0-30kV Local or Remote Programming
  • 0-300µA Local or Remote Programming
  • Polarity Reversible Upon Command in <1 Sec at No Load
  • Low Stored Energy, Current Limited Output
  • Full Feature Front Panel, Ideal for Laboratory Useage

Certifications & Compliance

Get a Quote

Auto-Reversing High Voltage Power Supply

Spellman’s CZE1000R is a full feature rack mountable high voltage power supply ideal for laboratory usage. It’s designed to meet the needs of applications requiring a hot switched reversible output voltage. The output polarity can be quickly and safely reversed via a front panel switch.

Both the output voltage and current are fully adjustable from 0 to 30kV and 0 to 300uA via front panel ten turn locking counting dials. Remote control operation is done by 0 to +10Vdc programming signals; either user generat­ed or using the provided +10 Vdc reference and external potentiometers.

Front panel voltage and current meters provide local monitoring. Voltage and current test points are provided such that 0 to 10Vdc corresponds to 0 to 100% rated output. A two position, normally closed, external interlock is provided for protection of external high voltage accessi­ble areas. If the interlock is opened the high voltage will shut off and fall to zero in less than one second and not be able to be re-energized until the interlock is closed.

Excellent load and line regulation specifications along with outstanding stability and low ripple of the CZE1000R assure a stable high voltage output for consistent process results.

  • Specifications
  • How to Order
  • Downloads

Specifications

  • Input Voltage:
    115Vac, ±10%, 50/60Hz
  • Input Current:
    Less than 1 amp
  • Output Voltage:
    0 to 30kV
  • Polarity:
    Auto reversible via front panel switch
  • Output Current:
    0 to 300µA
  • Power:
    9 watts, maximum
  • Line Regulation:
    0.01% for a 10% input voltage change
    Load Regulation:
    0.01% for a full load change
  • Ripple:
    0.1% Vp-p
  • Stability:
    0.02% per 8 hours (after 1/2 hr warmup)
  • Temperature Coefficient:
    100ppm/°C
  • Weight:
    22lbs. (10kg)
  • Dimensions:
    5.25”H x 19”W x 17”D (13.3cm x 48.3cm x 43.2cm).
  • Regulatory Approvals: Designed to meet EEC EMC Directive. Designed to meet EEC Low Voltage Directive.

How to Order

How to Order:
Standard: PART NO.: CZE1000R

Typical applications:

  • Electrospinning
  • Mass Spectrometry
  • Capillary Electrophoresis
  • Electrostatic Research
2026 High Voltage Reference Manual

We compiled all of our useful FAQ’s, application notes, articles, technical papers and three glossaries into a single handy reference guide.

Download
Service Solution Brochure

Spellman Repair Centers, Extended Warranty Programs, Preventative Maintenance Programs, Calibration Services and more.

Download